Title: When measuring soil water content, field practices affect neutron moisture meter accuracy
Abstract: Field practices for measuring soil water content with neutron moisture meters were assessed. One finding: Using meters of the same manufacturer with different components (such as different types of detector tubes) or meters of different manufacturers can greatly alter performance and affect results. Guidelines were developed for making standard counts and for making counts in the soil.