Title: A compact and portable X-ray beam position monitor using Medipix3
Abstract: The present work reports on the design and implementation of a novel portable X-ray beam diagnostics (XBPM) device. The device is transparent to the X-ray beam and provides real-time measurements of beam position, intensity, and size. The measurement principle is based on a pinhole camera which records scattered radiation from a Kapton foil which is placed in the beam path. The use of hybrid detectors (Medipix3) that feature a virtually noiseless readout system with capability of single photon detection and energy resolving power enables the diagnostics with a better resolution and higher sensitivity compared to the use of traditional indirect X-ray detection schemes.
Publication Year: 2014
Publication Date: 2014-12-16
Language: en
Type: article
Indexed In: ['crossref']
Access and Citation
Cited By Count: 3