Title: Radiation exposure to medical staff in interventional and cardiac radiology.
Abstract:The aim of this work has been to determine typical occupational dose levels in interventional radiology and cardiology installations and to relate doses to patient and occupational dosimetry through t...The aim of this work has been to determine typical occupational dose levels in interventional radiology and cardiology installations and to relate doses to patient and occupational dosimetry through the dose-area product. An experimental correlation between environmental dosimetric records and dose-area products in the centres studied was established. The study covered a sample of 83 procedures performed by 10 specialists in six laboratories. The radiologists and cardiologists monitored wore nine thermoluminescent chips next to eyes, forehead, neck, hands, left shoulder, left forearm and left arm during each single procedure. In addition, direct reading electronic devices for environmental dosimetry were placed in the C-arm of the X-ray system, to estimate roughly the occupational radiation risk level. Typical shoulder doses derived from electronic dosimetry range between 300 and 500 muSv per procedure, assuming no lead protective screens were used. Using these values and patient dose-area data from two laboratories, averaged ratios of 84 and 120 muSv per 1000 cGy cm2 are obtained for cardiology procedures. Finally, occupational dose reductions of approximately 20% when using highly filtered X-ray beams with automatic tube potential (kV) reduction (available in some facilities), and by a factor of about three when using ceiling mounted screens, have been found.Read More
Publication Year: 1998
Publication Date: 1998-09-01
Language: en
Type: article
Indexed In: ['crossref', 'pubmed']
Access and Citation
Cited By Count: 350
AI Researcher Chatbot
Get quick answers to your questions about the article from our AI researcher chatbot