Title: Automated system for inductance realization traceable to AC resistance with a three-voltmeter method
Abstract: An automatic system for the comparison of unlike impedances in the range 1 /spl Omega/-10 k/spl Omega/, based on the three-voltmeter method, has been constructed and is undergoing extensive tests as a system for realization of the inductance scale at the Istituto Elettrotecnico Nazionale Galileo Ferraris (IEN), in the range 100 /spl mu/H-10 H at frequencies of 120 Hz and 1 kHz. The main features of the system are a two-phase direct-digital synthesis generator for energizing the standards, complete four-terminal pair definition of the standards with automatic servosystems, the use of a single high-accuracy voltmeter and a switching unit, and optical links for the distribution of frequency references. Measurements at the metrological level have been performed at 1 kHz with inductors in the range 10 mH-1 H.
Publication Year: 2001
Publication Date: 2001-12-01
Language: en
Type: article
Indexed In: ['crossref']
Access and Citation
Cited By Count: 40