Title: Panasonic dosimetry system performance testing and results at nuclear accident dose levels 500 rad to 10,000 rad
Abstract: Panasonic thermoluminescent dosimeters (TLDs) are used as the photon dose assessment part of the personal nuclear accident dosimeter (PNAD) and may be used for the same purpose with the fixed nuclear accident dosimeter (FNAD). To demonstrate compliance with 10CFR835.1304 (and, its predecessor, DOE Order 5480.11), several sets of dosimeters were irradiated to photon doses above the upper limit of the DOELAP testing standard, DOE/EH-0026 and DOE/EH-0027. The upper range of the test was 10,000 rads, using both low energy (70 keV) and high energy (662 keV and 1,332 keV) sources. The testing indicated that the Panasonic TLD system is capable of meeting the requirements of 10CFR835.1304 and DOE Order 5480.11.