Title: Measurement of W-Values for Alpha Particles in Tissue Equivalent Gases
Abstract: W values (average energy loss per ion pair) were measured on four kinds of tissue equivalent gases for alpha-particles from 241Am by a modified Jesse method. A chart recorder following an electrometer figuring in Jesse's apparatus was replaced by a “peak holder” circuit and a pulse height analyzer, which made it possible to measure the pulse height distribution directly. The measured W values were 31.1 eV for Rossi's TE gas, 28.5 eV for Shonka's TE gas, 27.2 eV for propane-butane-air mixture and 27.1 eV for a new TE gas. Strickler's formula which has been used to predict W values of binary gas mixtures is modified into a formula applicable to the mixture of three gases or more. The estimated W values of Rossi's TE gas and propane-butane-air mixture from the modified formula were in excellent agreement with the measured results.
Publication Year: 1976
Publication Date: 1976-06-01
Language: en
Type: article
Indexed In: ['crossref', 'pubmed']
Access and Citation
Cited By Count: 10