Title: Development of a biochemical-pump for flow injection analysis system with chemiluminescence detection
Abstract: A biochemical-pump was constructed using a dialysis cell with catalase immobilized membrane, in which active pressure was produced by enzymatic reaction with hydrogen peroxide (H2O2) concentration. The biochemical-pump system provided non-pulsating flow. The flow injecton analysis (FIA) system was also fabricated using a photodiode sensor and the biochemical-pump with carrier solution including luminol and horseradish peroxidase (HRP), thus detection chemiluminescence (CL) indused by the injection H2O2.