Title: Apparatus for millinewton force measurements with millisecond response time
Abstract:Apparatus is described which is capable of measuring forces in the range 10-4-10-2 N with a response time of the order of a millisecond. The use of piezoresistive strain-gauge transducers extends the ...Apparatus is described which is capable of measuring forces in the range 10-4-10-2 N with a response time of the order of a millisecond. The use of piezoresistive strain-gauge transducers extends the response to DC. The noise level of less than 10-5 N is achieved with the aid of a simple but effective antivibration, screened enclosure. Application to dynamic meniscus rise measurements illustrates the capabilities of the system.Read More
Publication Year: 1979
Publication Date: 1979-01-01
Language: en
Type: article
Indexed In: ['crossref']