Title: Assessment of oxygen transfer in membrane oxygenators during clinical cardiopulmonary bypass
Abstract: Although functional replacement of the heart and lungs by a pump and oxygenator is a widespread surgical procedure, no widely accepted technique for describing gas exchange in oxygenators exists. In this study, 8 types of commercially available membrane oxygenator (2 flat sheet membrane, 4 gas in hollow fibre membrane and 2 blood in hollow fibre membrane) have been studied during clinical cardiopulmonary bypass. O2 transfer increased with blood flow rate but the O2 transfer at a given blood flow was lower than that obtained by the manufacturers in laboratory studies. Overall O2 transfer coefficients were calculated from the ratio of O2 transfer rate to an O2 difference expressed either as an O2 partial pressure or an O2 concentration. Specific O2 transfer coefficients (overall coefficient divided by membrane area) were similar for oxygenators with a flat sheet or gas in hollow fibre membrane configuration. The two types of oxygenator with blood in hollow fibre membranes had significantly lower (P less than 0.01) specific O2 transfer coefficients. This study shows that oxygenator gas transfer characteristics can be studied in the clinical environment and that O2 transfer coefficients can be related to oxygenator design features.
Publication Year: 1992
Publication Date: 1992-05-01
Language: en
Type: article
Indexed In: ['crossref', 'pubmed']
Access and Citation
Cited By Count: 8
AI Researcher Chatbot