Abstract: An inductively coupled surface coil for imaging the rat spine at 7 T is described. This planar circular probe was made from microwave substrate to limit the size of the coil and to minimize the magnetic susceptibility. The surface coil was used as a single transmit/receive coil and as a receive-only coil with a birdcage body coil for excitation. The signal-to-noise ratio (SNR) of the probe was compared to a 5-cm birdcage coil and exceeded the birdcage coil's SNR by three to six times at superficial structures. The main advantages of the probe are an improved SNR for superficial structures and a simple design and use. Images with 50 × 50 × 500 μm voxels were obtained of the rat spine with excellent anatomical detail.
Publication Year: 1992
Publication Date: 1992-01-01
Language: en
Type: article
Indexed In: ['crossref', 'pubmed']