Title: Digitale Röntgenthoraxaufnahmen mit einem Selendetektor: Prospektiver Vergleich mit einer konventionellen Film-Folien-Kombination
Abstract:To compare a new digital chest radiography system that uses amorphous selenium as the x-ray detector, with conventional radiography for the detection of pathological alterations of the chest.Two obser...To compare a new digital chest radiography system that uses amorphous selenium as the x-ray detector, with conventional radiography for the detection of pathological alterations of the chest.Two observers analysed pairs of posteroanterior and lateral chest radiographs of 95 patients. One pair of radiographs was obtained with the digital selenium chest radiography system, and the other with conventional film-screen technique. 9 criteria were rated using a 4 point scale. Technical parameters were standardised. Radiation dose was measured in both techniques and compared.A total of 855 criteria were rated. 740 findings were diagnosed in accordance on both techniques (740/855 = 87%). 115 criteria (115/855 = 13%) showed deviations. The mean radiation dose for the selenium detector was 0.02 cGy and for the conventional system 0.11 cGy.The exposure of radiation is lower using a selenium x-ray detector compared to conventional film-screen technique in chest radiography. The digital selenium system performs well in a clinical setting, providing visualisation of pathological findings as good as a standard screen-film system.Read More
Publication Year: 1997
Publication Date: 1997-02-01